Technical specifications

Technical data

Anode voltage 60–120 kV
Anode current 1–16 mA
Focal spot 0.5 mm, fixed anode
Image detector Flat panel
Image acquisition 200 / 360 degree rotation
Scan time 1–36 s
Typical reconstruction time 2–55 s

Maximum volume sizes

Volume size [cm]

Planmeca Viso G5

Maximum volume with a single scan:
Ø20 x 10

Maximum volume with multiple vertical scans:
Ø20 x 17

Planmeca Viso G5 maximum volume size

Planmeca Viso G7

Maximum volume with a single scan:
Ø30 x 20

Maximum volume with multiple vertical scans:
Ø30 x 30

Planmeca Viso G7 maximum volume size

Dental programmes

Volume size [cm]

Planmeca Viso G5

Tooth
Ø3 x 3 – Ø6 x 6

Teeth
Ø7 x 3 – Ø9 x 9

Jaw
Ø10 x 3 – Ø20 x 10

Face
Ø14 x 13 – Ø20 x 17

Planmeca Viso G7

Tooth
Ø3 x 3 – Ø6 x 6

Teeth
Ø7 x 3 – Ø12 x 10

Jaw
Ø13 x 3 – Ø17 x 17

Face
Ø14 x 12 – Ø30 x 20

Skull
Ø20 x 22 – Ø30 x 30

ENT (Ear, Nose, Throat) programmes

Volume size [cm]

Planmeca Viso G5

Nose
Ø7 x 5 – 9 x 9

Sinus
Ø10 x 11 – 20 x 150

Middle ear
Ø3 x 3 – 6 x 6

Temporal bone
Ø7 x 5 – 9 x 9

Airways
Ø9 x 8 – 11 x 10

Vertebrae
 Ø9 x 8 – 11 x 10

Planmeca Viso G7

Nose
Ø7 x 5 – 12 x 10

Sinus
Ø10 x 10 – 17 x 20

Middle ear
Ø3 x 3 – 6 x 6

Temporal bone
Ø7 x 5 – 12 x 10

Airways
Ø8 x 8 – 10 x 14

Vertebrae
Ø8 x 8 – 10 x 14

Physical space requirements

Width 155 cm (61 in.)
Depth 152 cm (60 in.),
Height* 240 cm (95 in.)
Weight 165 kg (lbs 364 )

With cephalostat:
Width 216 cm (85 in.)
Depth 154 cm (61 in.)
Height*180 cm (95 in.)
Weight 180 kg (lbs 397)

*The maximum height of the unit can be adjusted for offices with limited ceiling space.

Example installation

Included in delivery

Planmeca 3D unit with 3D reconstruction server

Planmeca 3D unit with 3D reconstruction server

Minimum set up

Client workstation and database server

  • Planmeca Romexis 3D Explorer
  • Database server
  • Planmeca Romexis Image Database

The client workstation and database server can also be in separate computers.

Planmeca client workstation and database server

Additional equipment

Additional diagnostic workstations with different software configurations

Planmeca Romexis tools:

  • 3D Explorer
  • 3D Cross Sections module
  • 3D TMJ module
  • 3D Implant Planning module
  • DICOM module

The client workstation and database server can also be in separate computers.

Planmeca additional workstations with different software configurations