Technical specifications

Technical data

Generator Constant potential, microprocessor controlled, operating frequency 66 kHz
X-ray tube Toshiba D-041SB
Focal spot size 0.4 mm according to IEC 60336
Cone diameter 60 mm (2.36 in.), Rectangular 33 x 43 mm (1.30 x 1.69 in.)
Max. symmetrical radiation field Ø60 mm at SSD 200 mm, Ø60 mm at SSD 300 mm, according to IEC 806
Total filtration min. 2.5 mm Al equivalent at 70 kV according to IEC 60522
Inherent filtration 1 mm Al equivalent at 70 kV according to IEC 60522
Anode voltage 60, 63, 66, 70 kV, ±2 kV
Anode current 8, 7, 6, 5, 4, 3, 2 mA ±(5% + 0.2 mA)
Exposure times 0.01–2 sec. ±(5% + 0.001 sec.), 24 steps
SSD (Source-Skin Distance) Standard/Long 200 mm (8 in.)/300 mm (12 in.)
Mains voltage 100 V~/110-115 V~/220-240 V~, 50/60 Hz
Duty cycle 1:30, automatic control
Electrical classification Class I Type B
Weight total 29 kg (64 lbs), tube head with standard cone 4.2 kg (9.3 lbs), tube head with long cone 4.5 kg (10 lbs) 
Colour White (RAL 9016)

Dimensions

Planmeca ProX technical dimensions

Installation options

Planmeca ProX wall mount

Standard wall mount

Planmeca ProX ceiling mount

Ceiling mount

Planmeca ProX Single stud mount

Single stud mount

Planmeca ProX Dental unit mount

Dental unit mount

Planmeca ProX Mobile stand

Mobile stand

Planmeca ProX Floor column

Floor column